Functional and technical properties - HygienicMaster
OI/FEX300/FEX500-EN FEX300, FEX500 167
14.1.7 Mechanical properties
Parts that come into contact with fluid
Part Standard Option
Lining PFA from DN 3
(1/10“)
PEEK DN 1 ... 2
(1/25 ... 1/12“)
-
Signal and
grounding electrode
CrNi steel 1.4539
(AISI 904L)
CrNi steel 1.4571
(AISI 316Ti)
Hast. C-4 (2.4610)
Hast. B-3 (2.4600)
Titanium, tantalum,
Platinum-iridium
Gaskets (for Weld
stubs, threaded
connection, Tri-
Clamp, external
threads)
EPDM (Ethylene-
Propylene) with FDA
approval, silicone
with FDA approval
(CIP-resistant, no oils
or grease)
Silicone with FDA
approval (option, oil
or grease resistant)
PTFE with FDA
approval (DN 3 ... 8
(1/10 ... 5/16“))
Gasket for 1/8"
sanitary connectors
PTFE Viton (only in
combination with
PVC process
connection)
Process connection
(welded stubs, Tri-
Clamp, etc.)
CrNi steel 1.4404
(AISI 316L)
-
Parts that do not come into contact with fluid
Standard Option
Flange
CrNi steel 1.4571
(AISI 316Ti)
-
Flowmeter sensor housing
Standard Option
Housing
Deep-drawn housing
CrNi steel 1.4301 (AISI 304),
1.4308
–
Terminal box CrNi steel 1.4308 (AISI 304)
–
Meter tube Stainless steel
–
PG- cable gland Polyamide, stainless steel
(version for ambient
temperature -40 °C (40 °F))
–
Change from one to two colu mns