134 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
15 Appendix
Overview of setting parameters and technical design
Measuring point:
TAG no.:
Sensor model:
Transmitter type:
Order no.:
Device no.:
Order no.:
Measuring medium temperature:
Voltage supply:
Lining:
Electrodes:
Exciter frequency:
C
zero
:
C
Span
:
System zero point:
Parameter
Adjustment range
Prog. Protection code:
................
0-9999 (0 = factory setting)
Language:
................
e.g., German, English, French, etc.
Flowmeter sensor
................
see name plate or the submenu ‘System Data/Sensor
Nominal size:
................
DN 1 to DN 1000
Q
max
:
................
0.05 Q
max
DN to 1 Q
max
DN
Pulse factor:
................
Imp./phys. Unit
Pulse width:
................
0.100 to 2000 ms
Offset suppression:
................
0 to 10 % of upper range value
Damping:
................
0.2 (0.07) to 20 seconds
Noise reduction
................
OFF / 1 / 2/ 3 / 4
Density:
................
0.01 g/cm
3
to 5.0 g/cm
3
Unit Q
max
.:
................
e.g., l/s, l/min, l/h, hl/s, hl/min, hl/h, etc.
Unit totalizer:
................
e.g. l, hl, m
3
, igal, gal etc.
Max. alarm:
................
%
Min. alarm:
................
%
Switching output:
................
Max. Alarm, Min. Alarm, Max./Min. Alarm, General alarm, empty conduit etc.
Contact input:
................
External zero return, Totalizer reset, External system zero, no function
Current output:
................
0/4
to 20 mA, 0/2 to 10 mA, 0 to 5 mA, 0 to 10-20 mA, 4 to 12 to 20 mA
I
out
with alarm:
................
0 %, 103 %, 3.8 mA, Low, High
Detector e. pipe:
................
ON / OFF
Calibrate e. pipe:
................
0 to 10000
Alarm e. pipe:
................
ON / OFF
I
out
for e. pipe:
................
0 %, 103 %, 3.8 mA, Low, High
Totalizer function:
................
Standard, difference totalizer
1. display line:
................
Q (%), Q (unit), Q (mA), counter V/R, TAG number, blank line, bar graph
2. display line:
................
Q (%), Q (unit), Q (mA), counter V/R, TAG number, blank line, bar graph
3. display line:
................
Q (%), Q (unit), Q (mA), counter V/R, TAG number, blank line, bar graph
4. display line:
................
Q (%), Q (unit), Q (mA), counter V/R, TAG number, blank line, bar graph
Operating mode:
................
Standard / Piston Pump / Fast
Flow direction:
................
Forward / reverse flow, Forward flow
Directional display:
................
Standard, inverse
Pulse output:
Active
Passive
Communication:
HART
PROFIBUS PA
FOUNDATION Fieldbus
Without
Diagnosis
Threshold monitoring
Error (E)
Warning (W)
Measured value
E/W
Min.
Max.
Measured value
E/W
Min.
Max.
Coil current AC
Coil voltage DC
Coil resistance
Coil temperature
Coil insulation resistance
DAC value
Electrode voltage
Electrode balance
Signal-to-noise ratio
Magnetic linearity