FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 129
Sensor SE21 / SE21F
Minimum permissible absolute pressure
Operation
Operation
1)
1) For CIP/SIP cleaning, higher temperatures are permitted for limited time
periods; refer to the table entitled ‘Maximum permissible cleaning
temperature’.
Maximum permissible cleaning temperature
max
max
amb.
If the ambient temperature is > 25 °C (77 °F), then the difference must be
subtracted from the max. cleaning temperature.
T
max – Δ °C, Δ °C = (Tamb - 25 °C)
Maximum Allowable Temperature Shock
Temp shock max. temp.
diff. °C
Maximum permissible ambient temperature as a
function of measuring medium temperature
Ambient temperature T
amb.
Measuring medium temperature T
medium
Figure 79: Temperature graph
Sensor material
Hast.-
C4
(2.4610)
(1.4539
[904 L]
for pipe
conn. and
Tri-
Hast.-
B3
(2.4600),
1.4539
(904 L),
1.4571
(316 Ti),
Titanium,
tantalum,
platinum
DN
⅛” sanitary connectors always with 2 grounding electrodes in
the same material as measuring electrodes, standard.
Process connection material
Stainless steel 1.4571 (316 Ti)
Stainless steel 1.4404 (316 L)
Stainless steel 1.4404 (316 L)
-Clamp
Stainless steel 1.4404 (316 L)
Stainless steel 1.4404 (316 L)
without/with
preamplifier, type A
with preamplifier,
type B
Stainless steel 1.4301 (304)
Aluminum alloy, painted, paint coat frame: dark
gray, RAL 7012 cover: light gray,
RAL 9002
Stainless steel 1.4301 (304)
Stainless steel 1.4301 (304)
Gasket material (internal)
-Clamp
-
Propylene) with FDA
approval, silicone with
FDA approval (CIP
-
resistant, no oils or
grease)
Silicon with FDA approval
(optional, resistant to
oils and grease)
PTFE with FDA approval
(DN
3 to 8)
Silicon (resistant to oil,
grease)