FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 101
Recommended settings for diagnosis limit values
The values indicated are only intended as a rough guide and may need to be adapted in line with on-site conditions.
These values depend on the flowrate, so no recommendations can be made here.
1)
Factory or commissioning fingerprint x 0.95
Factory or commissioning fingerprint x 1.05
This value depends on the measuring medium temperature T
medium
T
medium
T
medium
Factory fingerprint x 0.81
Factory fingerprint x 0.89
Factory fingerprint x 0.86
Factory fingerprint x 0.95
Factory fingerprint x 0.9
Factory fingerprint x 0.95
Factory fingerprint x 1.05
Factory fingerprint x 1.05
Factory fingerprint x 1.16
Factory fingerprint x 1.14
Factory fingerprint x 1.26
Factory fingerprint x 1.24
Factory fingerprint x 1.37
180 °C (356 °F) Factory fingerprint x 1.33 180 °C (356 °F) Factory fingerprint x 1.47
This value depends on the measuring medium temperature T
medium
T
medium
T
medium
−40 °C (-40 °F) Factory fingerprint * 0.71 −40 °C (-40 °F) Factory fingerprint * 0.79
Factory fingerprint * 0.81
Factory fingerprint * 0.89
0 °C (32 °F) Factory fingerprint * 0.9 0 °C (32 °F) Factory fingerprint * 1
Factory fingerprint * 0.95
Factory fingerprint * 1.05
60 °C (140 °F) Factory fingerprint * 1.19 60 °C (140 °F) Factory fingerprint * 1.31
Factory fingerprint * 1.28
Factory fingerprint * 1.42
130 °C (266 °F) Factory fingerprint * 1.43 130 °C (266 °F) Factory fingerprint * 1.58
180 °C (356 °F) Factory fingerprint * 1.62 180 °C (356 °F) Factory fingerprint * 1.79
This value depends on the ambient / measuring medium temperature, so no recommendations can be made here.
1) This function is only guaranteed if an electrode zero point adjustment has been carried out (see page 94).