102 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 8 Operation
… Additional information regarding use of enhanced diagnostic functions
Minimum value Maximum value
10 MΩ 99.9 MΩ
100
Design level of the flowmeter
DN 3 to10 (1/10 to3/8”) 300 300
DN 20 to25 (3/4 to1”) 400 400
DN 40 to50 (1 1/2 to2”) 500 500
DN 80 (3“) 500 300
DN 150 (6“) 400 600
DN 350 to400 (14 to16”) 500 700
This value depends on the measuring medium temperature T
medium
T
medium
T
medium
Factory fingerprint x 0.86
Factory fingerprint x 0.95
−20 °C (-4 °F) Factory fingerprint x 0.90 −20 °C (-4 °F) Factory fingerprint x 1
Factory fingerprint x 0.93
Factory fingerprint x 1.03
20 °C (68 °F) Factory fingerprint x 0.95 20 °C (68 °F) Factory fingerprint x 1.05
Factory fingerprint x 1.10
90 °C (194 °F) Factory fingerprint x 1.05 90 °C (194 °F) Factory fingerprint x 1.16
Factory fingerprint x 1.09
Factory fingerprint x 1.21
180 °C (356 °F) Factory fingerprint x 1.14 180 °C (356 °F) Factory fingerprint x 1.26
-to-noise ratio) 60 % of the commissioning fingerprint or a higher value,
depending on the application.
100 %