30 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
6 Electrical connections
Preparing and routing the signal and magnetic coil cable
Preparing for sensor model SE41F
Cut to length and terminate both cables as shown.
Note
Use wire end sleeves.
Signal cable D173D025U01
Magnet coil cable D173D147U01
Preparing the transmitter side
Preparing the sensor side
Cable identifiers
1 Measuring potential, yellow
2 Reference, white
3 Signal cable, red
4 Signal cable, blue
5 SE clamp
1
Magnetic coil, black
2
Magnetic coil, black
3
SE clamp
Figure 28: Preparing the signal and magnetic coil cable sensor model SE21, SE21F
Note
The shielding may not touch (signal short circuit).