40 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
6 Electrical connections
Terminal assignment
Retrofit for model 10D1422: DN 3 to DN 1000 (¹⁄₁₀ to 40”); model 10DI1425 and 10DS3111A-E:
DN 500 to DN 1000 (20 to 40”)
S4 transmitter (wall mount housing)
Sensor
1
Functional ground (busbar)
2
Pulse output
1)
3
Switch input
1)
4
Switch output
1)
5
Current output
1)
6
Power supply
Low voltage: 100 to 230 V AC, terminals L, N,
Frequency: 47 Hz ≤ f ≤ 53 Hz; 50 Hz power supply
Frequency: 56 Hz ≤ f ≤ 64 Hz; 60 Hz power supply
7
Shielded signal cable: ABB order no. D173D025U01
or incorporate in existing wiring
8
Reference cable:
Only for connection to model 10D1422
Close terminals: 6,7 and hook switch S903
9
Magnetic coil supply:
Magnet coil supply via supply power
0
Terminal board: D685A1020U03
Figure 37: Sensor standard DN 3 to DN 1000 (1/10 to 40”), transmitter field mount housing
1) See the chapter ‘Connection examples for peripherals’ in the operating instruction and/or on the data sheet
Note
We recommend that shielded output cables be used with the shields connected to the functional ground at one end.