32 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 6 Electrical connections
Connecting the sensor
Connecting the signal and magnet coil cables
The flowmeter sensor is connected to the transmitter via the signal / magnet coil cables (part no. D173D025U01 / D173D147U01). The
coils of the flowmeter sensor are supplied with a field voltage by the transmitter over terminals M1/M3. Connect the cables to the
sensor in accordance with the following drawing, using a screwdriver with proper size and width.
Terminal box model SE21, SE21F
red
blue
yellow
white
black
SE clamp
signal cable
ground connection
magnetic coil cable
red
blue
yellow
white
black
SE clamp
signal cable
ground connection
magnetic coil cable
Figure 30: Connection to sensor
Wires for the measuring signal (red, blue)
Shielding for signal wires
Power supply for preamplifier via signal cable shielding
Wire for reference signal (white)
Shielding for reference signal cable
Measuring ground (yellow)
Connections for magnetic field excitation (black)