FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 103
Displaying the diagnostic values
Measured values
A maximum of four diagnostic values can be displayed (Fprt1, Fprt2, Fprt3, Fprt4). These values are selected in the submenu
Diagnosis / Fingerprint / Fprt1 - 4. In addition to the type of measured value, the desired data set should also be set. This
combination of settings can be used to display all stored values.
To enable users to display the different numerical values, an ID code is displayed with each value.
As a result, each line appears as follows:
e.g., CV_RCL 80.5 ohm to display the current coil resistance.
CV _ RCL 80.5 ohm
< 100 X.X
RIS: Insulation resistance
< 100 X.X
UE1: Electrode voltage E1
UE2: Electrode voltage E2
SNR: Signal-to-noise ratio
DF_: Commissioning fingerprint
Product 1 fingerprint
P2_: Product 2 fingerprint