FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 103
Displaying the diagnostic values
Measured values
A maximum of four diagnostic values can be displayed (Fprt1, Fprt2, Fprt3, Fprt4). These values are selected in the submenu
Diagnosis / Fingerprint / Fprt1 - 4. In addition to the type of measured value, the desired data set should also be set. This
combination of settings can be used to display all stored values.
To enable users to display the different numerical values, an ID code is displayed with each value.
As a result, each line appears as follows:
e.g., CV_RCL 80.5 ohm to display the current coil resistance.
CV _ RCL 80.5 ohm
Data set
Type
Numerical value
Unit
CV_: Current value
UAC: Coil voltage (AC)
X.X
V
H1_: History 1
IAC: Coil current (AC)
X
mA
H2_: History 2
RCL: Coil resistance
Value > 100 X
Value
< 100 X.X
Ohm
H3_: History 3
TCL: Coil temperature
X
°C
H4_: History 4
RIS: Insulation resistance
Value > 100 X
Value
< 100 X.X
MOhm
H5_: History 5
UE1: Electrode voltage E1
X
µV
H6_: History 6
UE2: Electrode voltage E2
X
µV
H7_: History 7
BAL: Electrode balance
E1 ≥ E2 → +X
E1 < E2 → -X
%
%
H8_: History 8
LIN: Magnetic linearity
X
%
H9_: History 9
SNR: Signal-to-noise ratio
X
%
FV_: Default fingerprint
DAC: Maximum DAC value
X
Dig = Digits
DF_: Commissioning fingerprint
P1_:
Product 1 fingerprint
P2_: Product 2 fingerprint