94 FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
… 8 Operation
… Parameter descriptions
Submenu Enter the length of the signal cable.
Notice: Information must be entered to calculate the coil
resistance / coil temperature.
0 ... 200
0.0
Submenu Offset correction for temperature display.
-100 ... 100
0.0
Import RT20 Import the reference resistance for the coil.
See page 97 in this regard.
Notice: This menu is displayed only when the coil resistance
or coil temperature are enabled as an error or warning
message.
Yes, exit C/CE
New 500 ohms
Yes → Enter
Electr. Zero Point This adjustment is required in order to use the diagnostic
measurement values for the electrode voltage E1/E2 and the
electrode balance as well as the wiring errors for the electrode
state.
Electr. Zero
Point
Yes, exit C/CE Display the current (new) and stored (old) zero values for
electrode E1/E2.
Select ‘Enter’ to import the current zero points
Notice: The liquid in the sensor must be at absolute standstill.
Displays stored electrode zero and current electrode AC
voltage acc. to amount and phase.
Where
N1: Current measured electrode AC voltage E1
N2: Current measured electrode AC voltage E2
A1: Stored zero point E1
A2: Stored zero point E2
Yes, exit C/CE