FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 95
Submenu / parameter
Value range
Remark
Submenu
Status
Error register
After calling up the error register, the error register text is
displayed with the number of pending errors. Press ENTER
again to display the pending errors with error number and
text. If you press ENTER again, you will be asked whether you
want to reset the error or view the help text. Select STEP to
display all of the pending errors with error numbers and text.
Press ENTER to reset the pending errors.
Amount: 1
C (Current)
Ext. Database
Reset: Enter
Help text: Step
0
Empty pipe
Warning register
After calling up the warning register, the warning register text
is displayed with the number of pending warnings. Press
ENTER again to display the pending warnings with warning
number and text. If you press ENTER again, you will be asked
whether you want to reset the warnings or view the help text.
Select STEP to display all of the pending warnings with
warning numbers and text. Press ENTER to reset the pending
warnings.
Amount: 2
3 (Pending)
* Test Mode *
Reset: Enter
Help text: Step
2
Counter reset
Power outage
Display the power outages during operation.
0
HMx ----------
Status register for enhanced diagnostics.
For more information see page 97.
HMi ----------
EMx ----------
EMi ----------