FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E 63
Submenu / parameter Value range Remark
Submenu
This menu does not appear for the PROFIBUS or FOUNDATION
fieldbus.
Data Iout
Adjust Iout 4 mA View only Display adjustment value for transmitter for the current output
at 4.0 mA.
4.0 mA
Adjust Iout 4 mA
View only
Display adjustment value for transmitter for the current output
at 20.0 mA.
20.0 mA
Submenu
Flowmeter sensor
Sensor model
Display the connected flowmeter model: The selection is
imported with the FRAM automatically by the transmitter. If an
older flowmeter sensor model (see list) is selected, the FRAM
with Cs = 100%, Cz = 100% must be used. The models 10D1422
(10DI1425, 10DS3111A-C ≥ DN 500) can only be connected to a
specially designed terminal board on the transmitter. Close
switch S903 on the connection board. The other models are put
into operation with their assigned FRAM. For this reason, when
older model flowmeter sensors are connected to the transmitter,
a ‘Primary Setup’ menu is displayed at initial startup. Additional
settings can be entered in the system data submenu ‘Primary’.
See also page 109.
DS2_, DS4
Line frequency View only Display the mains frequency for which the flowmeter sensor was
calibrated. Displayed only for 50/60 Hz older models. Not
displayed, if SE41F or SE21 sensor is selected.
50 Hz