8-4
Questions? Refer to Page 11-22 For Technical Support Instructions.
Strip Lot
• Decide whether an operator can continue to test (“allow test” or “warn”) or not (“lockout”) when a strip lot is scanned or
entered to the meter for a lot of strips that is not on the approved strip lot list. The maximum number of strip lots that can
be downloaded to the FreeStyle Precision Pro meter is 36: 18 Glucose and 18 Ketone.
Other
• Decide what additional information to download to a meter through the data management system: lists of operators, lists
of strip lots, or lists of patients.
• Determine whether the meter displays results in mmol/L or mg/dL (Glucose only).
• Decide whether any additional free text comments are stored with a test record. These can be required, optional or not
used.
• Decide whether to configure the meter for proficiency tests.
• Decide whether to configure the meter for linearity tests.
• Determine what language (such as English) displays on the meter.
• Turn the audible indicator off/on.
• Determine whether to use alkaline, lithium or rechargeable batteries in the meter.
• Decide how soon the meter automatically turns off (4 - 10 minutes) after a patient test.
• Decide whether to enable ketone testing.