EasyManua.ls Logo

Abbott FreeStyle Precision Pro - Troubleshooting Out-of-Range Control Test Results

Abbott FreeStyle Precision Pro
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11-2
Questions? Refer to Page 11-22 For Technical Support Instructions.
Reasons Ketone Results May Be Higher Than Expected:
• Haematocrit is lower than the acceptable limit for the test strips, as indicated on test strip package insert.
• Serum or plasma samples were used instead of whole blood.
Reasons Ketone Results May Be Lower Than Expected:
• Haematocrit is higher than the acceptable limit for the test strips, as indicated on test strip package insert.
• Water or alcohol remaining on the puncture site.
If test results appear higher or lower than expected for reasons not described above, please repeat the test using a new test strip.
If the results still appear higher or lower than expected, contact your local Abbott Diabetes Care Inc. oce or distributor.
Troubleshooting Out-of-Range Control Test Results
Repeat the test for that control solution and make sure that the operator meets the conditions in this checklist:
• Eliminate any air bubbles in the control bottles tip.
• Wipe the control solution nozzle with a clean gauze or tissue before and after each test. Liquid left on the tip from previous
tests may have a glucose or ketone concentration higher than expected.
• Calibrate the meter using the barcode for the test strip used.
• Scan or enter the correct 5-digit lot number for the control solutions.
• Confirm that control solutions and test strips have been stored within the ranges specified on their respective cartons and
package inserts.
• Check that the temperature conditions in the room where the tests are being performed are within the acceptable
operating range as specified in the test strip package insert.
Chapter 11: Troubleshooting

Table of Contents

Related product manuals