Chapter 3: Patient Test
Patient Test Procedure - Precision Xceed Pro Glucose
Use the following procedure to perform a patient test using the Precision Xceed Pro Blood Glucose Test Strip.
What You See on the Display What You Do Comments
1. Press On/Off to turn on the meter.
The Abbott logo screen will appear in
black for a few seconds and then will
appear in grey to ensure that the display is
functioning properly. Next, the software
version will briefly appear.
2. Press 1 to select Patient Test. The meter starts in Test Mode. The Menu
button will toggle the meter to Menu
Mode.
Note: The following section illustrates some common settings. Depending on your facilitys specific settings,
some screens will display differently or not at all. If the meter does not display the screens shown,
please refer to Chapter 11, Troubleshooting.