Chapter 1: Overview
Control Solutions
Control solutions are used to perform regular quality control checks on the meter to ensure it is functioning correctly. Control
testing is further described in Chapter 4. Use MediSense Glucose and Ketone Control Solutions with the FreeStyle Precision Pro
system. ese control solutions are available in one, two or three level congurations.
Calibration Verification Controls
Use RNA Medical brand Glucose and β-Ketone Calibration Verication Controls to conrm the calibration and analytical
measurement range of the meter for FreeStyle Precision Pro Blood Glucose or Blood β-Ketone Test Strips. is is called linearity
testing and requires a kit with 5 levels of control solution. Linearity testing is further described in Chapter 7.
Replacement Strip Port Module
A replacement strip port module is available if the system becomes disabled as a result of blood, uid or other contaminant
entering the strip port. Repeated strip port error messages may also indicate the need for replacement. e strip port module
replacement procedure is described in Chapter 12.
Data Management System
Abbotts FreeStyle PrecisionWeb Point-of-Care Health Management System (PrecisionWeb) is necessary to upload results from
the FreeStyle Precision Pro meter. Data upload from remote locations requires either terminal servers or networked PCs running
Abbotts Data Repeater soware. Data can also be uploaded through a wireless network when the individual meters and data
management system are properly enabled. Wireless conguration details are described in the optional Wireless Set-Up Utility
CD.