Warnings
12 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
2 Warnings, Cautions, and Symbol Denitions
System Messages
Respond appropriately to all system messages. If the message cannot be cleared,
contact an ACIST representative.
Safe Use of Equipment
No modication of equipment is allowed.
e CVi system may only be interfaced with X-ray equipment that is
certied to be in compliance with IEC 60601-1, third edition.
To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth ground.
To avoid the potential for electric shock to the patient, X-ray equipment
used in conjunction with the CVi system must provide two means of patient
protection.
To avoid the potential for electric shock to the patient, do not touch the
patient while making connections to the injector head or to the power
supply.
To avoid the potential for electric shock, do not touch the patient and the
X-ray interface connector pins simultaneously.
To meet the debrillator protection specied, the CVi system must be used
only with accessories (including transducers and adapter cables) that are
specied by ACIST Medical Systems.
Never disconnect a CVi system from the mains power source while the
main power switch is on. Make sure the main power switch is turned o
before disconnecting.
e CVi system should not be used in rooms containing magnetic
resonance imaging (MRI) equipment.