Warnings
12 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
2 Warnings, Cautions, and Symbol Denitions
System Messages
Respond appropriately to all system messages. If the message cannot be cleared,
contact an ACIST representative.
Safe Use of Equipment
• No modication of equipment is allowed.
• e CVi system may only be interfaced with X-ray equipment that is
certied to be in compliance with IEC 60601-1, third edition.
• To avoid the risk of electric shock, this equipment must only be connected
to a supply mains with protective earth ground.
• To avoid the potential for electric shock to the patient, X-ray equipment
used in conjunction with the CVi system must provide two means of patient
protection.
• To avoid the potential for electric shock to the patient, do not touch the
patient while making connections to the injector head or to the power
supply.
• To avoid the potential for electric shock, do not touch the patient and the
X-ray interface connector pins simultaneously.
• To meet the debrillator protection specied, the CVi system must be used
only with accessories (including transducers and adapter cables) that are
specied by ACIST Medical Systems.
• Never disconnect a CVi system from the mains power source while the
main power switch is on. Make sure the main power switch is turned o
before disconnecting.
• e CVi system should not be used in rooms containing magnetic
resonance imaging (MRI) equipment.