ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 45
Perform Contrast Purge5 Perform an Injection
X-Ray Sync (Optional)
e X-RAY SYNC button toggles between sync modes.
ON OFF
Contrast injection may be initiated
from either the hand controller or
the X-ray imaging system.
Contrast injection is controlled
by the CVi system using the hand
controller.
Notes
• Injections initiated by an X-ray imaging system always use xed injection rate
mode.
• The hand controller may be disabled with some X-ray imaging systems.
• X-ray imaging may be initiated by the CVi system with some X-ray imaging
systems.
• X-Ray Sync is available only with peripheral mode injections.
1. In the SELECT MODE box, press PURGE.
2. Press SYRINGE or TUBING to select the component to purge.
3. When purging the syringe, follow these steps:
a. Press PURGE.
b. e system automatically purges air from the syringe.
c. Press OK.
4. When purging the tubing, follow these steps:
a. Turn the stopcock o to the patient.
b. Press OK.
c. Press and hold PURGE.
d. e system purges air as long as the PURGE button is held, to a
maximum of 10 seconds.
e. Press CANCEL to return to normal operation.
1. In the SELECT MODE box, press SALINE.
2. Press and hold FLUSH. e tubing will be ushed while the FLUSH
button is held, to a maximum of 10 seconds.
Flush Saline Using the Hand Controller
Instead of using the control panel, a saline ush can be initiated using the hand
controller.
1. Press the saline (S) button on the hand controller. e CVi system will
begin a 10 second ush of the saline components.
2. To cancel the saline ush, press the saline (S) button on the hand
controller a second time.
Perform Contrast
Purge
Perform Saline
Flush using 0.9%
Normal Saline