End the Case
48 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
6 End a Case
4. Disconnect and discard the patient tubing, saline tubing and bag, the
hand controller and the manifold assembly.
5. Press OK.
Note
Patient kits must be removed before proceeding to the next step.
6. Remove a disinfecting cap from the strip included with the A2000 syringe
kit.
7. To release the alcohol contained in the disinfecting cap, push and twist
the cap onto the end of the syringe’s male luer connector until the cap will
turn no further. e disinfecting cap must remain rmly in place on the
male luer for a minimum of 5 minutes to ensure proper disinfection.
8. Soware version 7 or higher only: Press DONE aer fully disinfected.
9. Leave the disinfecting cap in place until connecting a new manifold to the
syringe.
Important
A new disinfecting cap must be placed on the male luer after each use.
10. Press CARDIAC or PERIPHERAL to select the procedure mode.
11. When ready for the next case, continue the setup procedure at Load
Contrast, using the existing syringe assembly.
If Using a New Syringe
1. On the touchscreen, press END CASE, then press OK to conrm.
2. Select NO when prompted to shut down. e system displays the number
of cases for which the syringe has been used.
3. Press YES to start a new case with a new syringe.
4. Close the white slide clamp on the contrast container tubing.
5. Remove and discard the patient tubing, saline tubing and bag, hand
controller, and the manifold assembly.
6. Press OK.
7. Wait until the syringe plunger has disengaged, then open the syringe
chamber.
8. Remove and discard the syringe.
9. Close the syringe chamber.
10. Press RESTART.
11. Follow the instructions beginning with Load the Syringe Assembly on
page 34.