Change Color Scheme7 Supplementary Procedures
e color scheme of the buttons on the control panel can be changed by
pressing COLOR. Toggling between Fixed rate mode and Variable rate mode
always changes the color scheme.
Press SYSTEM INFO to cycle through CVi system data:
• HW Conguration – Board and control panel IDs
• SW Conguration – System soware and rmware versions
• Current Session Info – Case number and patient manifold valve (pValve)
failures
• System Info – Total number of power-ups and cases run since memory was
cleared
Note
The KVO function is available only with peripheral mode injections.
KVO Injection
A KVO injection periodically injects a pulse of saline to keep the uid pathway
open to the patient during delays between injections.
To initiate a KVO injection, press KVO.
KVO Rate
e default ow rate for KVO injections is 1.0 ml/min. Press KVO RATE to
select a dierent ow rate for the KVO function. e available range is
0.1 ml/min–10 ml/min.
Change Color
Scheme
System Info
Keep Vessel Open
(KVO) Injection
(Optional)