EasyManua.ls Logo

ACIST CVi - Ending the Final Case of the Day; Handling Emergency Cases

ACIST CVi
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 49
End the Case6 End a Case
If Final Case
1. On the touchscreen, press END CASE, then press OK.
2. Select YES when prompted to shut down. Follow the screen prompts.
3. Close the white slide clamp below the contrast container.
4. Remove and discard the AT P54/AT P65 Hand Controller and patient
tubing, the BT2000 manifold assembly, contrast container, and the saline
bag.
5. Wait until the syringe plunger has disengaged, then open the syringe
chamber.
6. Remove and discard the A2000 syringe.
7. Discard any remaining contrast media and saline in accordance with your
institutions procedures.
8. Turn o the power switch on the power supply.
9. Refer to Chapter 7 Supplementary Procedures for detailed instructions on
decontamination and daily cleaning.
Emergency Case
Important
It is preferred to use the A2000 syringe for ve patient procedures.
However, the system will allow one additional use to accommodate an
emergency case. After a syringe has been used in six patient procedures, the
system requires that a new syringe be used. The system software will prevent
a user from beginning a new case until the syringe has been replaced. The
A2000Syringe Kit must be discarded after six cases. See If Using a New Syringe
on page 48 to begin a new case with a new syringe.
1. On the touchscreen, press END CASE, then press OK. Select NO when
prompted to shut down.
2. Press NO to start a new case with the current syringe. e system displays
the message below in the status area of the main screen.
3. Press YES to use a new syringe and refer to If Using a New Syringe on
page 48. Press NO to use the syringe for one additional emergency case.