9 Specications
ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 81
Fill 1–2 ml/s
Syringe air purge 6 ml/s.
Automated xed ow rate of 1.6 ml/s.
Maximum saline injection volume of 16ml per saline injection activation.
Cardiac procedure mode
Injection Type
Flow
(ml/s)
Volume
(ml)
Pressure Rise
Time
(s)
psi kPa
LCA 0.8–10.0 0.8–20.0 200–1200 1379–8274 0.0–1.0
RCA 0.8–10.0 0.8–20.0 200–1200 1379–8274 0.0–1.0
LV / Ao 0.8–40.0* 0.8–99.9 200–1200 1379–8274 0.0–1.0
Other 0.8–40.0* 0.8–99.9 200–1200 1379–8274 0.0–1.0
Peripheral procedure mode
Injection Type
Flow
(ml/s)
Volume
(ml)
Pressure Rise
Time
(s)
psi kPa
Pigtail 0.8–40.0* 0.8–99.9 200–1200 1379–8274 0.0–1.0
Selective 0.8–15.0* 0.8–99.9 200–1200 1379–8274 0.0–1.0
Microcatheter 0.8–3.0 0.8–10.0 200–300 1379–2068 0.0–1.0
Other 0.8–40.0* 0.8–99.9 200–1200 1379–8274 0.0–1.0
* When using 5Fr catheters, ow rate reduction may occur with > 15ml/sec
ow rate settings and Pressure Limit may occur with ow rate settings
> 20ml/sec. When using 4Fr catheters, ow rate reduction may occur with
> 12ml/sec ow rate settings and Pressure Limit may occur with ow rate
settings > 15ml/sec.
Load Rate
Saline Rate
Contrast
Injection
Parameters
Specications