ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 85
Hemodynamic Transducer9 Specications
Refer to the following specications for Smiths Medical LogiCal® Pressure
Transducer product code MX960. e following specications are current as of
the date on this document and are subject to change by the manufacturer.
Specication Range
Pressure measurement –30 mmHg to +300 mmHg
Sensitivity 5 µV / V / mmHg nominal
Output impedance
300 Ohm ± 10%
Input impedance 630 Ohm nominal
Balance
0 mmHg ± 50 mmHg
Temperature sensitivity coecient
0 ± 0.1% / °C
Temperature coecient of calibration
value
0 ± 0.3 mmHg / °C
Excitation 4–8 VDC to 5 kHz
Risk current <5 µA at 115 VAC, 60 Hz
Overpressure withstand –400 mmHg to +4000 mmHg
Debrillation capacity 5 discharges in 5 min at 360 J
15% pressure transducer bandwidth > 200 Hz
e CVi system synchronizes with the following X-ray imaging systems. is
table reects current interface capabilities as of this revision. Please contact the
manufacturer for any updates if your system is not identied here.
Manufacturer Model
GE Discovery Series
Innova Series
OEC 9x00 Series
Optima Series
Unity Series
Phillips Allura Xper Series
Shimadzu AngioSpeed Series
Bransist Sare Series
Digitex Sare Series
Heart Speed Series
Siemens AXIOM Series
Toshiba Innix Cleve I Series
Innix i-series
Hemodynamic
Transducer
Supported
Imaging Systems