Monthly Inspection
54 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
7 Supplementary Procedures
8. Open the syringe chamber using the syringe door pin. Apply DisCide
ULTRA to the accessible syringe chamber surfaces and in particular, the
metal syringe nozzle channel and the syringe door pin. Allow the material
to remain moistened for a minimum of 1 minute prior to removal per the
user instructions. Slide the syringe chamber back into position
9. Apply DisCide ULTRA to the contrast sensor making sure both sides of
the sensor are fully covered. Allow the material to remain moistened for a
minimum of 1 minute prior to removal per the user instructions.
10. Remove the backlight cover by applying light pressure to the top or
bottom of the cover from the backside. Apply DisCide ULTRA to the
cover making sure both sides fully covered. Allow the material to remain
moistened for a minimum of 1 minute prior to removal per the user
instructions.
11. Open the peristaltic pump by using the pump door handle. Apply DisCide
ULTRA to the peristaltic pump (saline pump) making sure both the pump
door handle and retaining door latch are fully covered. Allow the material
to remain moistened for a minimum of 1minute prior to removal per the
user instructions. Close the pump door aer cleaning/disinfecting.
12. Apply DisCide ULTRA to the control panel making sure both sides of the
control panel are fully covered. Allow the material to remain moistened
for a minimum of 1 minute prior to removal per the user instructions.
Inspect the following on a monthly basis:
• Inspect all cables for cuts, cracks, or worn insulation, as well as separation of
cables and connectors. Replace damaged cables.
• Check for poor contact on all connectors.
• Check the syringe chamber for cracks, opacity, scratches, and other damage.
Replace if damaged.
• Check to make sure that the bed rail mount (if applicable) is securely
fastened.
• Check for wheel damage on the CVi system cart (if applicable).
For optimal performance, an ACIST Medical Systems representative should
perform routine annual maintenance.
When storing a cart-mounted system, wrap or attach the power cord to the
cart to avoid accidents caused by loose cords. Lock the wheels to prevent
unintentional movement of the system.
Monthly
Inspection
Annual
Preventive
Maintenance
Storage of
Cart-Mounted
Systems