6 End a Case
ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 47
Once a case is completed, end the case and follow one of the procedures below,
depending on whether you will
Begin a new case with the same syringe.
Use a new syringe for a new case.
Shut down the system.
End Case
1. Disarm the system from inject mode by pressing CANCEL in the system
message area.
2. Record all relevant case information such as contrast delivered and last
injection.
Note
Last Injection indicates the volume and highest ow rate achieved during
the most recent injection.
Important
All case data are cleared when the case is ended.
If Beginning a New Case with the Same Syringe
Important
It is preferred to use the A2000 syringe for ve patient procedures.
However, the system will allow one additional use to accommodate an
emergency case. After a syringe has been used in six patient procedures, the
system requires that a new syringe be used. The system software will prevent
a user from beginning a new case until the syringe has been replaced. The
A2000Syringe Kit must be discarded after six cases. See If Using a New Syringe
on page 48 to begin a new case with a new syringe.
1. On the touchscreen, press END CASE, then press OK. Select NO when
prompted to shut down. Follow the screen prompts.
2. e system displays the number of cases for which the syringe has been
used. Press NO to start a new case with the current syringe.
3. Close the white slide clamp on the contrast container tubing.
End the Case
End a Case