7 Supplementary Procedures
ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 51
Regular cleaning helps prevent contrast buildup, which can interfere with the
operation of the system. When cleaning the system, observe the following
guidelines:
• Turn o the power switch on the power supply before cleaning.
• Only use DisCide® ULTRA wipes and/or DisCide® ULTRA spray (Palermo
Healthcare) as the cleaning/disinfection agent.
• Do not use detergents or other cleaning agents.
• Do not spray water directly on the CVi system.
• Do not immerse any component in water. Do not allow water to drip onto
the injector head, control panel, power supply, or cables.
• Do not use sharp objects.
Warning
A disinfecting cap must be properly placed onto the multi-use syringe
before cleaning.
Aer every patient procedure check to see if any contrast or other uids spilled
on the CVi system. If so, use the following procedure to clean and disinfect the
CVi system.
General Cleaning
Instructions
Supplementary Procedures