End the Case
50 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
6 End a Case
4. Close the white slide clamp on the contrast container tubing.
5. Disconnect and discard the patient tubing, saline tubing and bag, the
hand controller and the manifold assembly.
6. Press OK.
Note
Patient kits must be removed before proceeding to the next step.
7. Remove a disinfecting cap from the strip included with the A2000 syringe
kit.
8. To release the alcohol contained in the disinfecting cap, push and twist
the cap onto the end of the syringes male luer connector until the cap will
turn no further. e disinfecting cap must remain rmly in place on the
male luer for a minimum of 5 minutes to ensure proper disinfection.
9. Soware version 7 or higher only: Press DONE aer fully disinfected.
10. Leave the disinfecting cap in place until connecting a new manifold to the
syringe.
Important
A new disinfecting cap must be placed on the male luer after each use.
11. Press CARDIAC or PERIPHERAL to select the procedure mode.
12. Open a new manifold kit.
13. Using aseptic technique, remove the disinfecting caps and connect the
manifold to the syringe.
14. Clip the manifold into the manifold valve sensor.
15. Make sure the white slide clamp on the contrast container tubing is open.
Note
If the white slide clamp is closed, the system may display Manifold Valve
Open or Syringe Valve Open indicators or messages.
16. Continue the setup procedure at Load Saline Tubing Assembly on
page 37.