EasyManua.ls Logo

ACIST CVi - Cleaning and Disinfection Procedure

ACIST CVi
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
General Cleaning Instructions
52 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
7 Supplementary Procedures
Cleaning / Disinfection Procedure
If the CVi system needs to be cleaned of biohazards such as blood or other
bodily uids, ACIST recommends using DisCide ULTRA as a cleaning
and disinfection agent. Wear protective gloves. Discard all materials used
to decontaminate the system in accordance with all local, state, and federal
regulations, codes, and directives.
CAUTION
To avoid equipment damage:
Do not apply bleach or other cleaners except as specied. Damage and
degradation of the system and its parts can result if such solutions are
applied.
Do not apply cleaning solutions to the rubber membrane of the
pressure transducer backplate. Applying cleaning solutions to the
rubber backplate can degrade the transducer, resulting in poor pressure
waveforms.
Perform the following steps to clean and decontaminate the CVi system:
1. Turn o the power switch on the power supply before cleaning.
2. Remove all single-use patient kits from the equipment that was used in
the previous case, if applicable.
3. Remove all excess material such as contrast, blood, or saline from the
surfaces of the CVi injector using the DisCide ULTRA wipes. Use DisCide
ULTRA spray and a so bristle brush, as needed, to help remove any built
up areas.
Notes:
Do not use non-specied cleaning/disinfection agents or detergents.
Do not use sharp objects in the cleaning/disinfection.
4. Visually inspect the area that has been cleaned for any remaining soil or
material. If there is visible soil, repeat the cleaning steps until all visible
soil has been fully removed.

Table of Contents

Other manuals for ACIST CVi

Related product manuals