EasyManua.ls Logo

ACIST CVi - Hemodynamic Monitoring Troubleshooting

ACIST CVi
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACIST | CVi® System User's Guide 901418-003,01 2019-08 English (USA) | 71
Troubleshooting Hemodynamic Issues8 Troubleshooting
Zero Fluctuations
Was zero set up at midaxillary?
ӽ An elevation dierence between the transducer and stopcock can cause
an oset of 7 mmHg to 10 mmHg (0.9 kPa to 1.3 kPa).
ӽ Zero with saline in the patient tubing, the stopcock open to air, and the
stopcock and transducer located at midaxillary.
Did the zero change aer a large injection?
ӽ Aer a large injection, use a hand syringe lled with saline to ush
contrast from the patient catheter. en re-zero the transducer.
ӽ Re-zero the transducer before performing critical measurements.
ӽ Be sure to zero at midaxillary.
Did the zero change aer a saline ush?
ӽ An elevation dierence between the transducer and stopcock can cause
an oset of 7 mmHg to 10 mmHg (0.9 kPa to 1.3 kPa).
ӽ Zero with saline in the patient tubing, the stopcock open to air, and the
stopcock and transducer located at midaxillary.
ӽ Ensure that there is saline and/or contrast in the patient tubing.
ӽ Manually ush the patient catheter with a hand syringe and re-zero the
transducer when performing critical measurements.
If a second transducer was added aer the initial zero, re-zero the
transducer. Ensure that the stopcock is open to air, and that the stopcock
and transducer are located at midaxillary.
Check the transducer for damage or dislodgement. If the transducer needs
to be replaced, disconnect from the patient, replace the manifold/transducer
kit, remove all air from the system, and resume the case.
Check the transducer backplate:
ӽ e backplate is rated for 500 cases.
ӽ Make sure that the transducer is seated properly on the backplate.
ӽ Check the backplate connector.
ӽ Replace the backplate if it is worn out, cracked, or damaged. When
inspecting the backplate, pay special attention to the condition of the
membrane and to the rails that hold the transducer in place.
Troubleshooting
Hemodynamic
Issues

Table of Contents

Other manuals for ACIST CVi

Related product manuals