Frequently Asked Questions
66 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
8 Troubleshooting
What can be done about air in the patient tubing before or beyond
the air column detect sensor?
1. Turn the stopcock so that the patient is disconnected from ow.
2. Press PURGE and then press TUBING.
3. Purge the patient tubing until all air has been removed.
Note
The air column detect sensor is not a substitute for user vigilance.
The saline tubing is being moved by the saline pump. How can I
stop it?
Lower the rear tubing guide so that it stops the saline tubing from advancing
without crimping the saline tubing.
Use only saline tubing from ACIST Medical Systems. e performance of
other tubing cannot be guaranteed.
Why is the hand controller not working?
Check for kinks in the hand controller tubing.
Make sure that the luer connections on the hand controller are securely
fastened to the control panel. e control panel can be rotated up for a
better view of the luer connections. If the hand controller is disconnected,
reconnect and recalibrate by pressing CALIBRATE HC on the control panel.
e hand controller must be replaced. Connect a new hand controller and
recalibrate by pressing CALIBRATE HC on the control panel.
The saline button on the hand controller is not working. Why?
Make sure that the luer connections on the hand controller are securely
fastened to the control panel. e control panel can be rotated up for a
better view of the luer connections.
If needed, reconnect the hand controller. Two to three turns are needed to
make a good connection. If the hand controller is reconnected, calibrate it
again.
Make sure that the saline pump is closed and the rear tubing guide is
lowered onto the saline tubing.
Make sure that the saline tubing is not bunched up in the pump.
Make sure that the stopcock is open.
The control panel touch screen is not responding. Why?
e CVi system may be in Standby mode. Check whether the Standby button
has been pressed. If it has, press the Standby button again, and then press OK
on the control panel.