EasyManua.ls Logo

ACIST CVi - Page 90

ACIST CVi
94 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EMC Requirements
90 | English (USA) 2019-08 901418-003,01 ACIST | CVi® System User's Guide
10 EMC Tables
Recommended Separation Distances for the CVi System
e CVi system is intended for use in the electromagnetic environment in
which radiated disturbances are controlled. e customer or user of the
CVi system can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications
equipment and the CVi system as recommended below, according to the
maximum output power of the communications equipment.
Service Band
(MHz)
Maximum Power
(W)
Separation
(m)
TETRA400
380 – 390 1.8 0.3
GMRS460,
FRS460
430 – 470 2,0 0.3
LTEband13,17
704 – 787 0.2 0.3
GSM800/900,
TETRA800,
iDEN820,
CDMA850,
LTEband5
800 – 960 2,0 0.3
GSM1800,
CDMA1900,
GSM1900,
DECT,
LTEband1,3,4,25
UMTS
1 700 – 1 990 2,0 0.3
Bluetooth,
WLAN802.11b/g/n
RFID2450,
LTEband7
2 400 – 2 570 2,0 0.3
WLAN802.11a/n
5 100 – 5 800 0.2 0.3

Table of Contents

Other manuals for ACIST CVi