Accessories and options 137
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
PEEPhigh
4 to 30 mbar **
FOT operator buttons
Next step
It activates the next step and can be used to skip the remaining
time of the stabilization period.
ON /OFF
It starts the whole FOT procedure and can be used to terminate it
anytime.
Repeat
It can be used to repeat the reactance measurement.
Reverse
It can be used to skip the upcoming higher-pressure level
measurements and continue at the previous step.
Feedback information fields automatically calculated values
PIPmax
maximum 100 mbar in HFOV and 80 mbar in Conventional mode
Step
minimum 1 mbar
Xrs
500 to 100 mbar / Lps
* If FOTsteps is set to 1 then high-pressure setting is hidden and Pmean low and PEEPlow text is
changed to Pmean and PEEP respectively.
** The difference between the high and low-pressure parameters can be minimum 1 mbar) or more if
the setting of the number of FOT measurements (FOTsteps) is bigger than 3 because of the step
size cannot be lower than 1 mbar.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals