74 Alarms
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
CPAP [DUOPAP-mode] lower limit: 2 mbar below set CPAP value
7.1.2 Configurable alarms
Following alarm limits are displayed and can be configured in the Alarm Limits menu.
High SpO
2
alarm: 2 to 99%, off
Low SpO
2
alarm: off, 1 to 98%
High Pulse Rate alarm: 35 to 235 bpm, off
Low Pulse Rate alarm: off, 30 to 230 bpm
Low Perfusion Index alarm: off, 0.03 to 19%
Low SIQ Alarm: off, 5 to 100%.
7.2 Alarm log
Switch to Alarm Limit menu.
Press the button underneath Show Log
to access the alarm overview.