Ventilation modes 95
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
11.3 SIMV
Breaths triggered by the patient will be displayed in green (default settings).
Ventilation with specified pattern and
frequency synchronous to the patient’s
independent breathing.
The patient can spontaneously breathe
in between breaths but receives no
pressure support.
For weaning from ventilation.
If Apnea is detected, ventilation will
commence with the specified TI and TE
frequency.
The synchronization window for the next
mechanical breath is maximum ½ T
e
.