System operation 55
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
5.7 Emergency shutdown
WARNING
In the event of a fault resulting in loss of normal control thee mergency shutdown procedure
may be used. This procedure disconnects all power to the ventilator and any other devices
connected to the ventilator.
Before performing this procedure, ensure that the alternate ventilation method (for
example: manual resuscitation) is present, ready and standing by.
1. Remove patient from the ventilator.
2. Press and hold the ON / OFF button
(1) for one to two seconds. Then
release it.
3. After one to five seconds, press and
hold the ON / OFF button (1) again
for at least five
seconds.
4. After the device has
powered OFF, a notice
tone will sound.
5. Press the Alarm Silence button to confirm system switch OFF.
6. Hold the button for at least three seconds until the WARNING LED
stops.