138 Ventilator service and maintenance intervals
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
13 Ventilator service and maintenance intervals
Always clean and disinfect equipment or equipment components before any maintenance – including
when returning the equipment for repair.
The expected service life of fabian is 8 years, assuming the servicing operations are performed as
specified by the manufacturer.
Maintenance and Safety Inspections
Perform the following work:
• Check alarm and limit value functions
• Check Electrical connections
• Check Pressure connections
• Check Safety Shutdown Calibration
• Device check as specified by manufacturer
Replace the following components:
• Air and Oxygen Input Filters
• Cooling Air Filter (casing bottom)
• Exhalation membrane
• Flow sensor
• O
2
sensor
Replace the following components:
• Air and Oxygen Input Filter
• Battery Pack
• Calibration Valves
• Cooling Air Filter (casing bottom)
• Exhalation membrane
• Flow sensor
• Input pressure regulator pressurized Air /
Oxygen
• Proportional Valves pressurized Air /
Oxygen mixer
• Lithium Battery (real time clock battery)
• Membrane HFO module
• O
2
sensor
Maintenance and safety inspections must be performed by ACUTRONIC
Medical Systems trained personnel with access to suitable testing and
measuring equipment.
Maintenance Interval Indicator. The symbol will appear in the Information bar 30
days before the next maintenance has to be performed.
fabian must not be disposed of with household waste but must be handed in for
disposal as electrical and electronic equipment, including battery / batteries
according to applicable national, regional and local rules, laws and regulations.
Accessories and consumables must be disposed of in accordance with the
relevant instructions for use. For information, contact your local environmental or
regulatory agency, or an appropriate waste disposal company.