58 Configurations menu
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
6.1.1 Flow sensor
Flow sensors can be calibrated in the upper half of the Calibration menu.
To calibrate a Neonatal Flow sensor, select the Neonatal button.
To calibrate a Pediatric Flow sensor, select the Pediatric button.
The Calibration procedure is identical for both types of flow sensors and described below.
Ensure that gas will NOT flow through
the Flow sensor during Calibration.
The sensor can be held occluding either
one or both ends with a sterile glove to
ensure zero gas flow.
1. Press the Flow Cal button.