76 Alarms
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
7.4.1 Alarms table
This table is intended to help you determine and resolve the cause of an alarm message.
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
1.
Acoustic audio
alarm test
failed
System failure
Acoustic audio test has
failed
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
2.
Cooling fan
defect
System failure
Fan not moving
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
3.
COM interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
4.
DIO interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
5.
I2C interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
6.
parallel
interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
7.
SPI interface
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
1
8.
checksum
conductor PIC
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
2
9.
checksum
monitor PIC
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
2