30 System functions and displays
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
4.3.3 Ventilation parameter dependency
If parameters are mutually regulating, an
arrow will appear on the parameter requiring
modification indicationg the required direction.
In this example, P
insp
is being reduced but is
NOT able to be reduced further as it is limited
by the PEEP setting. The limiting value is
indicated by the arrow.
List of Dependencies:
O
2
Flush min. 2 Vol.% above O
2
setting, max. 100 Vol.%
Only Volume limit or Volume guarantee possible
Pinsp min. 2 mbar above PEEP
Ppsv min. 2 mbar above PEEP
Pmax min. 2 mbar above PEEP
Pmanual min. 2 mbar above CPAP
Ppsv Pinsp
Pmanual min. 1 mbar above Pmean
Pmean rec min. 2 mbar above Pmean
Rise-Time ≤I-time
4.3.4 Locking ventilator parameters
Some Ventilator parameters are “Locked
to prevent unusual high values.
When this value is reached, a key symbol
appears and a “Notice Signal” will sound.
The high value must be confirmed by pressing
the rotary push knob again.
(see the following table)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals