158 Technical specifications
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
Resistance Range: 0 to 5000 mbar / L/s
Precision: ±8%
Resolution: 0.1 mbar / L/s
Overdistension Index C20 / C Range: 0 to 5
Resolution: 0.1
18.4 Resistance values
System Resistance at 30 Lpm <20 mbar / L/s
Inspiratory Resistance <12 mbar / L/s
Expiratory Resistance <8 mbar / L/s
18.5 Ventilation menu settings
Alarm delay
0 to 30 seconds
Automatic O
2
calibration
21% Vol.
21% Vol. and 100% Vol.
Eflow
1 to 32 Lpm
Manual breath in HFO
Enabled
Disabled
Max. time Manual breath
2 to 30 seconds
NIV Leak compensation
Off / Low / Middle / High
NIV Tube set
MediJet
Infant Flow (Infant Flow or
Inspire)
Infant Flow
LP
P
psv
parameter
Set as absolute value
Set as above PEEP parameter
Ratio of I-time
Set as frequency / Itime
Set as Itime / E-time
Termination criteria PSV
1 to 85%
Trigger (Conventional)
Volume trigger
Flow trigger
Pressure trigger
Unit for Pressure
mbar
cmH
2
O
Use BTB for Vt
Enabled
Disabled
V
limit
/ V
guarant
V
limit
V
guarant

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals