Sterilization / cleaning / disinfection 139
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
14 Sterilization / cleaning / disinfection
14.1 fabian HFO
The device must be prepared after each patient treatment.
WARNING:
The device must NOT be sterilized under any circumstances.
14.1.1 Preparation for disinfection
1. Switch off fabian before cleaning and unplug at the mains in order to prevent damage due to
penetration by liquid.
2. Remove any visible soiling on the fabian enclosure with the cleaning agent based on the option
selected at the next chapter below.
14.1.2 Manual disinfection options
Option A: CaviWipes Surface Disinfectant Wipes (EPA REG. NO.: 46781-14)
1. Unfold the CaviWipe and wipe the surface of the enclosure completely.
Note: The surface treated must remain visibly moist for three minutes.
Use more wipes if necessary to keep the surface moist for three minutes.
2. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one minute.
3. Allow to dry in the air.
Option B: 10% bleach (Clorox EPA REG. NO.: 5813-1)
1. Prepare a 10% bleach solution using purified water.
Use a clean lint-free cloth dampened with the bleach solution.
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
Note: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary to keep the surface moist for five minutes.
3. Wipe the enclosure with a clean lint-free cloth dampened with purified water for one minute.
4. Allow to dry in the air.
Option C: 5% H2O2, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H2O2 solution using purified water.
Use a clean lint-free cloth dampened with the H2O2 solution
2. Wipe the enclosure completely.
Remove any heavy soiling with additional dampened cloths if necessary.
Note: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary to keep the surface moist for five minutes.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals