26 System functions and displays
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
The alarms can be set between the following ranges:
(in HFO mode)
Lower limit
0 to 54
PIP [mbar]
(in DUOPAP mode)
Upper limit
4 to 30
PEEP [mbar]
(in DUOPAP mode)
Lower limit
5 to 19
CPAP [mbar]
(in NCPAP mode)
Upper limit
4 to 30
Lower limit
5 to 19
CPAP [mbar]
(in CPAP mode)
Upper limit
9 to 40
Lower limit
10 to 39
DCO
2
[mL
2
/ sec]
(in HFO mode)
Upper limit
2 to 10000, OFF
Lower limit
OFF, 1 to 9900
WARNING: The respiratory rate and minute volume monitors are calculated as averages
using an update period of 6 seconds. In the event that the monitored breath rate falls below
11 bpm, these values are not reported.
4.2.5 Graphics display
Displays the current Pressure-, Volume- or
Flow measurements as a diagram.
Use the Graphics key
to access the Graphics menu.
Three waveforms are simultaneously
displayed.
You can also switch to Loop Display view.
In this view two loops and one of three
waves are displayed.
LOOP Display:
Pressure / Volume
Volume / Flow