EasyManuals Logo

ACUTRONIC fabian HFO User Manual

ACUTRONIC fabian HFO
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #61 background imageLoading...
Page #61 background image
Configurations menu 61
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
Manual Calibration:
Press the 21% button to
perform One Point Calibration
at 21% O
2
.
Press the 21 + 100% button
to perform a Two Point
Calibration at 21% and 100%
O
2
.
After one of these buttons has been pressed, the
Calibration procedure starts. If the procedure has
been successful, a Green check mark will
appear.
The O
2
sensor can be turned OFF in the event
that the O
2
sensor reaches end of life during use.
You must then use a secondary respiratory gas
monitor until the ventilator can safely be taken
out of service to replace the O
2
sensor.
After a successful calibration, the date of the last calibration is displayed in the corresponding area.
Automatic Sensor Calibration runs every time the equipment restarts and every
24 hours. The gas concentration to patient is not altered during this calibration
procedure.
6.1.3 etCO
2
module
Please refer to chapter “etCO
2
monitoring”
6.1.4 SpO
2
module
Please refer to chapter “SpO
2
module”

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals