Sterilization / cleaning / disinfection 141
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
4. Allow to dry in the air.
Option C: 5% H2O2, Hydrogen Peroxide (EPA REG. NO.: 335-1)
1. Prepare a 5% H2O2 solution using purified water.
Use a clean lint-free cloth dampened with the H2O2 solution
2. Wipe the cable completely.
Remove any heavy soiling with additional dampened cloths if necessary.
Note: The surface treated must remain visibly moist for five minutes.
Use more dampened cloths if necessary to keep the surface moist for five minutes.
3. Wipe the cable with a clean lint-free cloth dampened with purified water for one minute.
4. Allow to dry in the air.
14.2.2 Flow sensor
WARNING:
Conduct hygienic preparation of flow sensor prior to each use.
14.2.2.1 Disassembly
Remove the flow sensor from the flow sensor cable.
14.2.2.2 Preparation for sterilization by automated cleaning
Please follow the steps below:
1. Rinse under running cold tap water (<35°C / 95°F) to remove excess soil.
2. Prepare a detergent bath using Neodisher Mediclean Forte solution at the manufacturer’s
recommendation of 10 mL per Liter using utility (tap) water 20°C / 68°F - 30°C / 86°F.
3. Transfer the flow sensor on to the 4-Level manifold rack accessory (or other appropriate rack
system) contained inside the washer for processing.
4. Select the appropriate cycle as listed below:
Phase
Recirculation Time
(Minutes)
Temperature
Detergent Type and Concentration
(If applicable)
Pre-wash 1
02:00
Cold tap water
N/A
Wash 1
02:00
45°C / 113°F
Tap water
(Set Point)
Neodisher Mediclean Forte 10 ml per
Liter
Neutralization
Wash
02:00
Warm tap water
Neodisher Z 1.0 ml per Liter
Rinse 1
01:00
Warm tap water
N/A
Dry Time
7:00
90°C / 194°F
(Set Point)
N/A