164 Electromagnetic compatibility statement
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
19 Electromagnetic compatibility statement
NOTE:
fabian HFO is a MEDICAL APPLIANCE subject to specific precautionary measures
regarding EMC and must be installed and started up according to the notices and the
instructions in this manual.
WARNING:
Portable medical HF communication devices could impact fabian HFO.
WARNING:
NEVER use fabian HFO directly adjacent to, or stack with other equipments. If
unavoidable, be sure to monitor the equipment for proper operation with this setup.
WARNING:
Using other accessories, cables or converters with fabian HFO could result in increased
electromagnetic interference and reduce the immunity of fabian HFO.
NOTE:
The key performance characteristics of the fabian HFO ventilator are:
fabian HFO must operate within the defined specifications and the medical purpose.
Failure to meet these specifications will result in respiration being stopped.
For this reason, a second stand-alone ventilation unit must always be available,
(for example: Resuscitation Bag).
19.1 Devices from SN AI-01500 and AL-00400
Guideline and Manufacturer Declaration Electromagnetic Emissions
The device fabian HFO is intended for operation in the environment described below. The customer or operator
of the fabian HFO apparatus should ensure it is operated in this type of environment.
EMI
Measurement
Compliance
Level
Electromagnetic Environment Guideline
RF Emission
CISPR 11
Group 1
The device fabian HFO uses RF Energy solely for internal operation. Its RF
emission is thus very low and interference with adjacent apparatuses is
unlikely.
RF Emission
CISPR 11
Class A
The emissions characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11 class B is normally required)
this equipment might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.”
harmonic
Distortion
IEC 61000-3-2
Class A

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals