88 Ventilation parameters
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
10 Ventilation parameters
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
I-Flow or Rise Time
Will allow for use of either I-flow or Rise time based
pressure rise control during inspiration
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
Rise Time
Adjustable rise time during inspiration
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
Inspiratory flow
Adjustable flow during inspiration
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
Expiratory flow (base flow)
The continuous expiratory flow is variable separate
from the inspiratory flow.
Flushing dead space in the Y-piece
Maintaining the PEEP
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
Positive end-expiratory pressure
Pressure at the airway connection port at the end of
expiration.
• SIPPV
• SIMV
• SIMV + PSV
• PSV
• duoPAP
Trigger
The trigger sensitivity can be adjusted from level 1 – 10.
This corresponds to 10% - 25% of the exhaled tidal
volume( Vte ) with volume trigger and 0.12–1.2 l/min in
10 steps (neonatal mode) / 0.6-6.0 l/min in 10 steps
(pediatric mode) with flow trigger.