Ventilation parameters 89
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
There are 3 trigger functions:
Volume trigger
Flow trigger
Pressure trigger
Which may be selected in the configurations
menu Ventilation
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• PSV
• duoPAP
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
• duoPAP
Frequency (Rate)
Frequency of mandatory breaths. In SIPPV mode the
delivered breath rate may be higher depending on the
patient’s spontaneous breathing.
Backup Rate in PSV and S-duoPAP modes
• IPPV
• SIPPV
• SIMV
• SIMV + PSV
Inspiratory pressure PSV
Inspiratory pressure during inspiration in PSV mode.
Inspiratory pressure
Inspiratory pressure during mandatory inspiration in
PSV mode.
Upper inspiratory pressure in duoPAP mode.
The ventilator provides continuous flow for producing
airway pressure at the set CPAP level.
at the set CPAP level
Continuous Positive Airway Pressure
Continuous flow for producing airway pressure
at the CPAP level
Manual Inspiratory pressure
Applying a manual breath
Next breath is possible after a block period of 200 ms.