42 System operation
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
5.1.3 Connect the tubing set
WARNING: ELECTRICAL HAZARD
NEVER use anti-static or electrically conductive tubes.
WARNING:
The pressure gradient of the fabian ventilator system measured at the patient connection
port can increase when accessories or other components are attached to the system.
Change in pressure gradient can adversely affect the performance of the fabian system.
1. Inspiratory Insp. port.
2. Expiratory Exp. port.
3. Proximal Pressure Prox. port.
4. Use FG Connection when connecting
to an INOvent Flow sensor.
1. Connect inspiratory tube to Insp. port.
2. Connect expiratory tube to Exp. port.
3. Connect pressure measuring tube to
Prox. port.
4. Connect HFO tube to HFO adapter.
Only required when operating in HFO
mode.
NOTE: Always hold tubes by the collar when connecting and disconnecting to prevent
damage.
The Expiration membrane can be easily
removed by turning the membrane holder
counter-clockwise.
Pay attention to the correct
installation position of the Expiration
Membrane.
The label, a 6-digit serial number (SN),
must be legible upon installation.
(See illustration).