Alarms 77
fabian HFO | SW 5.1.x
Ref: 113003.EN / Date 2020-02-10
#
Alarm Text
Alarm Type
Cause
Solution
Main
Priority
Sub
Priority
10.
Low physical
memory -
please reboot!
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
2
11.
Air supply
pressure
System alarms
Air supply pressure too
low.
Ensure pressure above
2 bar.
I
2
12.
Oxygen supply
pressure
System alarms
Oxygen supply
pressure too low.
Ensure pressure above
2 bar.
I
2
13.
Safety relay
defect
System failure
Safety relay defective.
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
2
14.
Tube
Occlusion
System alarms
Expiration tube
blocked/kinked
Inspiration tube
blocked/kinked
Excessive deviation
between inspiration
and expiration
pressure sensor.
Check ventilation
tubes.
Check pressure
sensor connection
(proximal).
If changing the
breathing circuit
does not resolve the
issue, then
disconnect the
device from AC
power and contact
your local Acutronic
Medical Systems
Distributor
I
2
15.
Voltage
monitoring
System failure
System error
Disconnect the device
from AC power and
contact your local
Acutronic Medical
Systems Distributor
I
2
16.
Battery defect
Electrical power
No power or too littel
power from battery.
Change battery,
Check fuse.
I
2
17.
Charge battery
(<15min)
Electrical power
Ventilator is not
connected to mains
supply.
Remaining battery
time < 15 min.
Switch to mains
supply.
Alarm can be
silenced with Alarm
Silence button to
continue in the
battery mode
I
2
18.
Flat battery!
Electrical power
Remaining battery time
< 1 min. Ventilation
can no longer continue
on battery power
Immediately switch to
Mains supply or
access alternative
ventilation
I
2
19.
Exhalation
calibration
System alarms
Exhalation calibration
failed
Change exhalation
membrane holder or
membrane.
I
2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian HFO and is the answer not in the manual?

ACUTRONIC fabian HFO Specifications

General IconGeneral
BrandACUTRONIC
Modelfabian HFO
CategoryMedical Equipment
LanguageEnglish

Related product manuals