95 17 Technical specifications
fabian Therapy evolution | SW V5.1.x
Ref: 121003.EN / Date: 26Jan2021
Acoustic Energy
201.9.6.2.1.101
TABLE: Acoustic energy
P
VBS used :
fabian Therapy evolution
Accessories :
Patient set
Humidifier (water level) :
N/A
Test lung :
Part No. 1116
Ventilator settings :
Acc. to TAB, 201.105 mid column
Measurement
No.
Location
Sound Pressure
Level
Remarks
Background
16
Lp A (dB)
1
2
33.9
1)
/50
Leq (dB)/max.@ (Hz)
1
4
32.8
2)
/50
Leq (dB)/max.@ (Hz)
25
6
36.3/12500
Leq (dB) max.@ (Hz)
25
8
33.1/12500
Leq (dB)/max.@ (Hz)
25
10
33.3/12500
Leq (dB)/max.@ (Hz)
25
12
37.6/12500
Leq (dB)/max.@ (Hz)
Calculated average
51.0/47.1
Lin/A weighted (dB)
Supplementary information:
LpA: A weighted sound pressure level in 1m distance: 40dB
1) Highest Level at Pos 2
2) Highest Level at Pos 4 etc.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ACUTRONIC fabian and is the answer not in the manual?

ACUTRONIC fabian Specifications

General IconGeneral
TypeNeonatal and Pediatric Ventilator
ManufacturerACUTRONIC Medical Systems AG
Modelfabian
Respiratory Rate1 - 150 bpm
FiO2 Range21 - 100%
HFOV Frequency5 - 20 Hz
Power Supply100-240 V AC, 50/60 Hz
Modes of VentilationSIMV, CPAP, HFOV
Tidal Volume Range2 to 300 ml
PEEP/CPAP0-20 cmH2O
Inspiratory Time0.1 to 3 seconds
Flow Rate2 - 30 L/min
Flow SensorIntegrated
Pressure Limit10-80 cmH2O